WP401C Industrial Pressure transmitter
This industrial pressure transmitter can be used to measure and control pressure for various industries, including petroleum and chemical industry, Electric power, water supply, Oil & Gas, environmental protection and other automatic control industries.
WP401C Industrial pressure transmitters adopt advanced imported sensor component, which is combined with the solid state integrated technological and isolate diaphragm technology.
The pressure transmitter is designed to work well under various conditions.
The temperature compensation resistance makes on the ceramic base, which is the excellent technology of the pressure transmitters. It has standard output signals 4-20mA, 0-5V, 1-5V, 0-10V, 4-20mA + HART. This pressure transmitter has strong anti-jamming and suits for long distance transmission application
Shell material:Aluminum Alloy
Wetted part material:SUS304(default material);SUS316
Special structure(noted when ordering)
Imported advanced sensor component
World-class pressure transmitter technology
Compact and strong structure design
Pressure Range can be adjusted in external
Suitable for all-weather harsh environments
Suitable for measuring a variety of corrosive medium
100% Linear meter, LCD or LED are configurable
Explosion-proof type: Ex iaIICT4, Ex dIICT6
Name | Industrial Pressure transmitter | ||
Model | WP401C | ||
Pressure range | 0—(± 0.1~±100)kPa, 0 — 50Pa~1200MPa | ||
Accuracy | 0.1%FS; 0.2%FS; 0.5 %FS | ||
Pressure type | Gauge pressure(G), Absolute pressure(A),Sealed pressure(S), Negative pressure (N). | ||
Process connection | G1/2”, M20*1.5, 1/2NPT, Customized | ||
Electrical connection | Terminal block M20x1.5 F | ||
Output signal | 4-20mA (1-5V); 4-20mA + HART;0-10mA(0-5V); 0-20mA(0-10V) | ||
Power supply | 24V DC; 220V AC, 50Hz | ||
Compensation temperature | -10~70℃ | ||
Operation temperature | -40~85℃ | ||
Explosion-proof | Intrinsically safe Ex iaIICT4; Flameproof safe Ex dIICT6 | ||
Material | Shell: Aluminium alloy | ||
Wetted part: SUS304 | |||
Media | Drinking water, waste water, gas, air, liquids, weak corrosive gas | ||
Indicator (local display ) | / | ||
Maximum pressure | Measurement upper limit | Overload | Long term stability |
<50kPa | 2~5 times | <0.5%FS/year | |
≥50kPa | 1.5~3 times | <0.2%FS/year | |
Note: When range <1kPa, only no corrosion or weak corrosive gas can be measured. | |||
For more information about this Industrial Pressure transmitter, please feel free to contact us. |